Root cause analysis of an aberrant thromboelastogram tracing – Lessons learnt


Department of Immunohematology and Blood Transfusion, Kasturba Medical College, Manipal Academy of Higher Education, Manipal, Karnataka, India

Date of Submission12-Dec-2019
Date of Decision27-Feb-2020
Date of Acceptance30-Mar-2020
Date of Web Publication17-Apr-2020

Correspondence Address:
Ganesh Mohan
Department of Immunohematology and Blood Transfusion, Kasturba Medical College, Manipal Academy of Higher Education, Manipal, Karnataka
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/GJTM.GJTM_73_19

Rights and Permissions
  Abstract 


Thromboelastography is a viscoelastic test which uses whole blood sample for routine coagulation screening purpose. Even though it is an easy procedure, many critical steps affect the final result of the test and we describe such an incident in this report. The root cause analysis of two aberrant TEG reports on two different occasions was initiated; the tracing initially was normal but developed an abnormal pattern toward the middle portion. The graph suddenly dropped to the baseline and from there, it was demonstrating an abnormal pattern. This can be due to power failure, vibrations during the test, cup dislodging from the pins during run, no validation prior to each test, and out of calibration. The root cause was identified as improper fixing of cups and pins leading to dislodging from the pins during the run and prevention of such incidents can be achieved by individual training of the staffs and periodical evaluation of their competency.

Keywords: Coagulation, hemostasis, quality assurance, root cause analysis, thrombelastography (TEG)


How to cite this article:
Mohan G, Shastry S, Kandasamy D, Prethika P A, Deepika C. Root cause analysis of an aberrant thromboelastogram tracing – Lessons learnt. Glob J Transfus Med 2020;5:80-3

How to cite this URL:
Mohan G, Shastry S, Kandasamy D, Prethika P A, Deepika C. Root cause analysis of an aberrant thromboelastogram tracing – Lessons learnt. Glob J Transfus Med [serial online] 2020 [cited 2020 Aug 11];5:80-3. Available from: http://www.gjtmonline.com/text.asp?2020/5/1/80/282749




  Introduction Top


Thrombelastogram (TEG® 5000, Hemonetics, USA) is a whole blood-based viscoelastic testing method which graphically depicts thein vivo hemostatic potential of the patient contributed by the enzymatic as well as cellular factors.[1] There has been a lot of modifications in the TEG technique to provide more reproducibility to the analysis. Of them, Citrated Kaolin with recalcification method is commonly used for routine purpose because of the test reproducibility and reagent stability.[2] We have developed age- and gender-matched normal reference ranges for our population [Table 1].[3] As with any other coagulation assays, various pre-analytical and critical analytical factors may influence the final TEG result[1] which needs to be addressed in the routine laboratory practice. Technicians should be given special training in sample collection, handling of the test consumables, and how to properly run the test, etc., before implementing TEG in the department. Technically, TEG is a simple test to perform, but often, we have overlooked many simple steps. Likewise, we present a report on how missing out small, simple steps might result in aberrant TEG results.


  Case Report Top


After a period of training and standardization of TEG, testing was started in our department, and on the 2nd month, we noticed an aberrant tracing during routine working hours. [Figure 1] represents aberrant TEG tracing observed on two separate occasions [Figures 1a, b and 2a, b] and depicts the TEG repeated on the same samples after immediate corrective action. There was no error message, and the “e-test” was validated by the technician before running the sample. The initial parameters of the tracing appeared normal [Figure 1]a and [Figure 1]b, but something went wrong while the test was going on. Confounding factors associated with such tracings have been identified as external vibrations during the test, improper fixing of cups and pins to machine resulting in dislodgment, and power failure during the tests.[1]
Table 1: Normal reference range for thromboelastography

Click here to view
Figure 1: (a and b) Aberrant TEG tracing

Click here to view


After a 360° observation and evaluation, we have identified the reason as improper fixing of cup and pin leading to dislodgment during the test. The manufacturer recommends that cups and pins have to be properly fixed by pressing onto the respective position and ensure it fits properly. Aberrant tracing appears when the cup gets dislodged from the pin as the impedance increases, and hence, the initial part will be normal.[2] Power failure and external vibrations will lead to abnormal tracing for that particular point of time, which were ruled out as a probable cause, as there was no interruption in power supply, and the machine was not disturbed during the run. Noncalibrated equipment might also lead to aberrant tracing, and in this case, calibration was done 2 months prior. When issues related to disposable material occur, an error message will appear on screen, indicating the user that cups and pins are not placed properly. We had two abnormal tracing without any error message and it may be due to the fact that the initial placement was satisfactory but loosely fixed onto the positions, thereby dislodging when clot strength resistance increases.[4]

Once the problem was identified, the current run was stopped and the test was repeated from the same sample obtained after informed consent from the patients, using fresh cups and pins, and the repeat run came as normal [Figure 2]a and [Figure 2]b. The root cause analysis was initiated and the results were issued after the corrective action. The deviation was documented in the TEG register and Ishikawa chart was plotted [Figure 3]. The root cause was identified as a technical error (man) and initiated the Plan–Do–Check–Act (PDCA cycle). A preventive action plan was charted, and the retraining of the staffs by the technical expert with specific attention given to individual technical person followed by their competency and skill evaluation was completed. Six months after implementing the preventive action, no such issues have been identified so far in the routine practice.
Figure 2: (a and b) Normal TEG tracing

Click here to view
Figure 3: Root cause analysis of TEG

Click here to view



  Discussion Top


Quality system essentials are one of the critical elements in sustaining and improving the quality of patient care. Deviations are said to happen in medical care, and various methods have been adopted to prevent the recurrence. PDCA cycle is one of the methods which helps us to continuously monitor the implemented corrective and preventive action and it is hassle free. The quality system ensures that a proper root cause analysis should be done whenever a deviation has been identified. Root cause analysis identifies the reason for such error or deviation, whereas our job is to prevent or minimize the deviation by implementing proper preventive action. Since ensuring quality is a dynamic evolving process and not a static one, continuous monitoring is required to ensure that the implemented actions are effective or not.

In our scenario, the observed deviation was cup and pin dislodgment during the TEG run without any error message. The other possible causes listed above were successfully ruled and the cause was a technical error of not appropriately fixing the cup and pin into the respective positions. The next thing was to implement preventive action, and hence, we conducted the training with individual attention and importance following each step, followed by an evaluation. We also did a PDCA cycle assessment and we have not observed such deviations after the RCA. This report highlights the importance of quality system in transfusion medicine and its role in delivering patient care. This report also brings light into the technical errors that may be seen with any investigation which should not be overlooked.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
MacDonald SG, Luddington RJ. Critical factors contributing to the thromboelastography trace. Semin Thromb Hemost 2010;36:712-22.  Back to cited text no. 1
    
2.
Nogami K. The utility of thromboelastography in inherited and acquired bleeding disorders. Br J Haematol 2016;174:503-14.  Back to cited text no. 2
    
3.
Ahammad J, Kurien A, Shastry S, Shah HH, Nayak D, Kamath A, et al. Age- and gender-related reference ranges for thromboelastography from a healthy Indian population. Int J Lab Hematol 2020;42:180-9.  Back to cited text no. 3
    
4.
Verma A, Hemlata. Thromboelastography as a novel viscoelastic method for hemostasis monitoring: Its methodology, applications, and constraints. Glob J Transfus Med 2017;2:124-9.  Back to cited text no. 4
    


    Figures

  [Figure 1], [Figure 2], [Figure 3]
 
 
    Tables

  [Table 1]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Case Report
Discussion
References
Article Figures
Article Tables

 Article Access Statistics
    Viewed262    
    Printed19    
    Emailed0    
    PDF Downloaded25    
    Comments [Add]    

Recommend this journal


[TAG2]
[TAG3]
[TAG4]